epMotion® 5075 TMX

epMotion® 5075 TMX
epMotion® 5075 TMX
play epMotion® 5075 TMX
play epMotion® 5075 TMX
play epMotion® 5075 TMX
play epMotion® 5075 TMX
play epMotion® 5075 TMX

Order Info

Product Information

The epMotion® 5075 TMX is the most flexible member of the epMotion family of automated pipetting systems. It includes an integrated TMX module to shake, heat or cool samples and reagents. The TMX module can be loaded and unloaded with the gripper and pipetting to the mixer is possible before and after mixing steps. The TMX module is fully software controlled enabling pipetting onto other worktable positions while the mixer is in operation.

Features

  • Like 5075 LH but with 11 ANSI/SBS microplate positions
  • Additional thermomixer
  • Temperature control range 15 °C below RT to 95 °C
  • Automatic fixing of labware for mixing
  • Pipetting at standard positions during shaking
  • Compact control panel, easy programming
  • Optional PC version
  • Optional software extensions for handling samples with barcodes or for integration to LIMS & ELN
  • Optional software extensions for integration to regulated process environments (GLP, GMP, 21 CFR)
  • Completely contained housing including door safety mechanism

Applications

  • Like 5075 LH plus
  • Bead applications with mixing and tempering
  • Sequencing and PCR clean-up
  • Nucleic acid purification
  • Cell disruption
  • Immunoassays
 epMotion® 5075 TMX
Gripper carrying capacity1,200 g
Mixing frequencyOff, 300 – 2000 rpm, optimized for labware and application
Mixing orbit3 mm
OperationElectronic
Pipetting typeAir-cushion
SLAS/ANSI deck positions11
Temperature homogeneity≤10 % on the block
Thermal module cooling down period of heating/cooling platefrom 25 to 4 °C in 5 minutes
Thermal module heating off period of heating/cooling platefrom 25 to 95 °C in 8 minutes
Thermal module setting range0-110 °C
Heating rate5 °C/min, on the heating plate
Cooling rate3 °C/min, on the cooling plate, above RT
Power supply100 – 240 V ±10 %, 50 – 60 Hz ±5 %
Max. power consumption105 W
Dimensions (W × D × H)107 × 61 × 67 cm
Weight w/o accessories116 kg
Temperature range-15 °C below RT to 95 °C, measured in labware at 75 % filling height
Conductor – X, Y, Z Positioning 
Random measurement error±0.1 mm
Systematic measurement error±0.3 mm
Detector 
Optical confocal infrared detectorContact-free recognition of fill level, inserted tools, labware, tip types and quantities
Dispensing tool 
Random measurement error (1 µL)1,2≤5 %
Random measurement error (1,000 µL)1,2≤0.15 %
Random measurement error (50 µL)1,2≤0.4 %
Systematic measurement error (1 µL)1,2±20 %
Systematic measurement error (1,000 µL)1,2±0.7 %
Systematic measurement error (50 µL)1,2±1.0 %
1in pipetting mode, free jet, without pre-wetting, with distilled water, at 20 °C
2See Eppendorf Application Note 168 for typical pipetting performance